EasyManuals Logo

Sartorius Cubis MCA2.7S User Manual

Sartorius Cubis MCA2.7S
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
Operating Instructions Cubis
®
MCA 99
Technical Data
Unit Value
Temperature
In operation °C +5 – +40
In operation for conformity-assessed devices:
See information on the device’s ID plate
During storage and transport °C -20 – +60
Relative humidity
At temperatures of up to 31°C % 80
Then linear decrease from 80% at 31°C to 50% at 40°C
No heat from heating systems or direct sunlight
No drafts from open windows, AC systems, or doors
No vibrations
No “heavy traffic” areas (personnel)
No electromagnetic fields
No dry air
15.3.2 Ambient Temperature for the isoCAL Function
Unit Value
Scope of application as per Directive 2014/31/EU
With isoCAL function °C +10 – +30
Without isoCAL function °C +17 – +27
15.3.3 Protection Class
IP Protection: Protected against dust and water
15.4 Materials
Housing: Die-cast aluminum, plastic PBT, Optiwhite float glass and stainless steel
1.4401 | 1.4404, PA handles, aluminum trim
Display and control unit: Plastic PBT, float glass and stainless steel 1.4301

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA2.7S and is the answer not in the manual?

Sartorius Cubis MCA2.7S Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA2.7S
CategoryScales
LanguageEnglish

Related product manuals