Technical Data
Unit Value
Temperature
In operation °C +5 – +40
In operation for conformity-assessed devices:
See information on the device’s ID plate
During storage and transport °C -20 – +60
Relative humidity
At temperatures of up to 31°C % 80
Then linear decrease from 80% at 31°C to 50% at 40°C
No heat from heating systems or direct sunlight
No drafts from open windows, AC systems, or doors
No vibrations
No “heavy traffic” areas (personnel)
No electromagnetic fields
No dry air
15.3.2 Ambient Temperature for the isoCAL Function
Unit Value
Scope of application as per Directive 2014/31/EU
With isoCAL function °C +10 – +30
Without isoCAL function °C +17 – +27
15.3.3 Protection Class
IP Protection: Protected against dust and water
15.4 Materials
Housing: Die-cast aluminum, plastic PBT, Optiwhite float glass and stainless steel
1.4401 | 1.4404, PA handles, aluminum trim
Display and control unit: Plastic PBT, float glass and stainless steel 1.4301