68 Operating Instructions Cubis
®
MCA
System Settings
Parameter Setting Values Explanation
Automatic output Off* Deactivates the automatic data output. The data output must
betriggered by means of an SBI command with “ESC P”.
Each weight value Starts the automatic data output with each value according to
thedisplay result.
1 second Starts the automatic data output every second.
2 seconds Starts the automatic data output every 2seconds.
5seconds Starts the automatic data output every 5seconds.
10seconds Starts the automatic data output every 10seconds.
30seconds Starts the automatic data output every 30seconds.
60seconds Starts the automatic data output every 60seconds.
Display SBI error
messages
Off* Displays no error messages.
On Displays error messages if errors occur when receiving
andexecuting SBI commands.
Formatting output
values
Output in one line with
full length
Outputs the elements in the full length, e.g., header and value.
The elements are separated by a space character, left-aligned,
and output without infilling with space characters.
Output in several lines
with 20 char. each
Distributes outputs, which due to their length do not fit in the
standard output format of 20 characters, over several lines with
20 characters each.
Aligns each line to the left and infills with space characters
tomake up 20 characters.
Cut off to a 20
character line*
Defines the output in a standard output format
of20characters.
Crops the end of longer outputs.
* Factory setting