58 Operating Instructions Cubis
®
MCA
System Settings
Parameter Setting Values Explanation
Stability delay Very short Sets the stability delay to “very short”: The stability symbol is
displayed after the stability criterion is reached.
Short* Sets the stability delay to “short”: The stability symbol only appears
after a short delay in order to provide a reliable result despite
fluctuations.
Medium Sets the stability delay to “medium”: The stability symbol only
appears after a longer delay in order to provide a reliable result
incase of higher fluctuations.
Long Sets the stability delay to “long”: The stability symbol only appears
after a long delay in order to balance out major instability.
Zeroing | taring Without stability Without stability: The function of the [Zero] or [Tare] key is
executed immediately once the key is pressed.
After stability* After stability: The function of the [Zero] or [Tare] key is only
executed after stability is achieved.
At stability At stability: The function of the [Zero] or [Tare] key is executed
ifstability exists when the key is pressed.
Automatic zeroing On* Activates automatic zeroing. The display is automatically set to zero
in case of a deviation of 0 less than (X).
Off Deactivates automatic zeroing. Zeroing must be triggered with
the [Zero] key.
Preset tare 1 On* Activates the second tare memory (preset tare).
Off Deactivates the second tare memory (preset tare).
Available units The availability of units may depend on national legislation and
istherefore country-specific. Multiple selection is possible.
mg – milligrams* The device displays the weight in milligrams.
g – grams The device displays the weight in grams.
ct – carats The device displays the weight in carats.
GN – grain The device displays the weight in grains.
dwt – pennyweight The device displays the weight in pennyweights.
mom – mommes The device displays the weight in mommes.
tol – tolas The device displays the weight in tolas.
MS – mesghals The device displays the weight in mesghals.
Available
resolutions
All digits on* “All digits on”: All digits are shown in the display. Not available on
conformity-assessed devices.
Last digit off at load
change
“Reduced by 1 digit for load change”: The last digit on the display
isswitched off until stability is achieved.
Last digit scale
interval 1
“Last digit scale interval 1”: The last digit always shows scale interval 1.
Last digit off “Last digit off”: The last digit is switched off.
*Factory setting for the “Standard Weighing” factory-set weighing profile