Interfaces
138 Cubis MSU User Manual
Configuring Network Operation (Ethernet)
Preparation: Before you can start with the configuration, you will require some information
about your network. Please contact your network administrator. The important
question is whether your network supports DHCP or not.
DHCP With DHCP, you can automatically link a (new) balance to an existing network
without having to configure it manually. Normally, the balance (client) only
requires the automatic acquisition setting for IP address (factory setting).
DHCP is supported In this case, no additional information is required.
DHCP is not supported In this case, you need the following information from the network administrator:
IP address
Subnet mask
Default gateway address
Device name (host name): This name can be changed if required.
This name is used to select the balance in the network.
Selecting Network Setting (Fixed Setting)
t If appropriate, toggle to the System Settings menu.
t Select the Device parameters submenu and slide the
right scroll bar down.
t Select Configure ports.
t Open the Network settings (Ethernet) submenu.
t To change the IP settings, select the input field behind
Network configuration.
t Select Use fixed IP address and confirm with OK.
t In the overview, check all settings and change as required.
Device name (host name): You can select this name freely.
This name is used to select the balance in the network.
t If you then want to establish an Ethernet connection, write down the IP
address now.
t To save the network settings, select Save.
y The “Configure ports” submenu is displayed.

Table of Contents

Other manuals for Sartorius Cubis Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis Series and is the answer not in the manual?

Sartorius Cubis Series Specifications

General IconGeneral
DisplayColor touch screen
Readability0.01 mg to 1 g
InterfacesUSB, Ethernet, RS-232
CalibrationInternal and external
HousingDie-cast aluminum
Response TimeVaries by model
Pan SizeVaries by model
Operating Temperature10°C to 30°C
ComplianceGLP
Power Supply100 – 240 V, 50/60 Hz
CapacityVaries by model (up to 64 kg)
RepeatabilityVaries by model
LinearityVaries by model