HF Series Generators
LV-DRAC
DR-1004R8
5
MINIMUM TIME FOR READY (DELAY)
MINIMUM TIME
3243SW1-x
M
I
N
I
M
U
M
T
I
M
E
(SECONDS)
1 2 3
0* ON ON ON
0.6 OFF ON ON
0.8 ON OFF ON
1.0 OFF OFF ON
1.2 ON ON OFF
1.5 OFF ON OFF
2.0 ON OFF OFF
3.0 OFF OFF OFF
* Note: It is recommended to set these switches at “0 seconds”.
FLUORO HOLD TIME
FLUORO HOLD TIME
3243SW2-x
F
L
U
O
R
O
H
O
L
D
T
I
M
E
(SECONDS)
1 2 3
0 ON ON ON
10 OFF ON ON
20 ON OFF ON
30 OFF OFF ON
40 ON ON OFF
60 OFF ON OFF
90 ON OFF OFF
120 OFF OFF OFF