HF Series Generators
T roubleshooting
TR-1005R4
62
ERROR CODE : E33
DESCRIPTION : No communication between Generator and Serial Console or PC Unit.
ERROR TYPE : Indicative.
APPLICABLE TO : All Generators with Serial Console.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Serial Communication Error.
POSIBLE CAUSES
Connecting cable between Console and Generator is loosen or damaged.
Damaged A TP Console CPU Board or Serial Console.
ACTIONS
1. Check connection cable between connector J7/J8 from A TP Console CPU Board to connector
J5 of Compatibility Module (as per Generator model) and then check cable from J5 to Serial
Console.
2. If error remains, check ATP Console CPU Board and Serial Console, replace defective part
found.
ERROR CODE : E34 -- TECHNIQUE ERROR
DESCRIPTION : Technique error.
ERROR TYPE : Informative without acoustic alarm or Fatal after exposition. It does not allow exposition.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment after parameter selection or after exposition.
INFORMATION / SYMPTOM : Indicated with warning “Technique error” in Consoles with written indicators. For the rest of
Consoles E34 appears in display .
POSIBLE CAUSES
The calibration for that parameters is wrong.
ACTIONS
1. If it occurs with time parameters close to 1mS, it means that the cable capacity is excessive for
a short exposition. Also the calibration for that parameters is wrong. Check Service Manual
Section: Exposure Time Adjustment.
2. If it happens at Fluoroscopy mode with ABC, it means that the equipment is not able to perform
that operation. Perform Jumper W1 in Fluoro CPU.
3. If it happens after exposition, it means a failure in exposition timer and backup has been
cut. Dangerous. Reset APR and reconfigure values.