HF Series Generators
T roubleshooting
TR-1005R4
63
ERROR CODE : E35 -- DOOR OPEN
DESCRIPTION : Door Open.
ERROR TYPE : Informative. It may inhibit Exposition depending on the configuration of SW1.2 at Console.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Indicated with warning “Door Open” in Consoles with written indicators. For the rest of Consoles
E35 appears in display. Does not allow “Prep” and “Exp”.
POSIBLE CAUSES
Door open or cable disconnected or cable not installed.
ACTIONS
1. Close door .
2. If that is not the reason for the error, check Jumper TS1-22 & TS1-23. See Installation Manual
Section: Door Interlock Signal.
ERROR CODE : E36 -- HEAT UNITS
DESCRIPTION : Heat Unit. Overheating.
ERROR TYPE : Informative without acoustic alarm. Does not allow expositions.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Indicated with warning “Heat Unit” in Consoles with written indicators. For the rest of Consoles
E36 appears in display. Does not allow “Prep” and “Exp”.
POSIBLE CAUSES
If it appears at initialization means that the thermostat of the selected tube is connected or it does not
exist, if so connect to ground the correspondent signal.
If it appears at any other moment is because the tube thermostat has been activated.
It also may appear during the calibration process due to the high number of starting.
ACTIONS
1. Check correct installation of signal Thermostat/Presostat. See Installation Manual. If tube has
not this signal, perform a Jumper indicated in Installation Jumper.
2. Wait until temperature is lowered.