HF Series Generators
Installation
IN-1052R0
8
For 80 kW Generators, the power supply line must be 480 VAC (or
530 VAC). If the Generator is supplied w ith an external step-up
autotransformer, it will include the cables to connect the autotransformer
to the Generator Cabinet. Power line s hould be connected to the
autotransformer terminals according to the line.
SEP-UP TRANSFORMER
Input: 380 VAC / 400 VAC / 415 VAC, Three-Phase.
Output: 480 VAC / 530 VAC, Three-Phase.
Max Power Output: 80 kW, 10% duty cicle
(approx. 5 seconds / minute)
Rising Temperature: 40
o
C
480 V
(or 530 V)
400 V
400 V
400 V
380 V
415 V
415 V
415 V
380 V
380 V
480 V
(or 530 V)
480 V
(or 530 V)
11. The whole System (Tables, Spot Film Devices, etc.) can be switched
ON/OFF when the Generator is switched ON/OFF. For this, power the
System through the output terminals of the Line Contactor 6K5 (upper
contactor close to the Input Transformer). These terminals are located
underneath the Line Contactor 6K5.
Illustration 2-4
Line Contactor 6K5
Line Contactor 6K5
Output Terminals
Input Transformer 6T2