108 FLOWSIC200 · Operating Instructions · 8013271/1CJ9/V2-0/2022-01 · © SICK Engineering GmbH
Malfunctions
6.2
Sender/receiver unit
Warning and error messages in SOPAS ET
Fig. 93 Directory “Diagnosis / Error Messages/Warnings“
Malfunctions listed below can probably be cleared onsite.
Description of possible causes and correction Display
SOPAS ET Possible cause Action
Communication A/B
Connection cable not connected correctly
Cables used do not comply with required specification
Both sender/receiver units are set to Master or Slave
A sender/receiver unit is defective
Check wiring (
p. 42, §3.3)
Parameter
Device has not yet been configured
Base parameters set to 0 after type change
Enter system data (again)
(
p. 67, §4.2)
Measuring range Configured measuring range overrun
Check parameter settings