Malfunctions
FLOWSIC200 · Operating Instructions · 8013271/1CJ9/V2-0/2022-01 · © SICK Engineering GmbH 109
6.3
Control unit
Malfunctions
Warning and error messages in SOPAS ET
Fig. 94 Directory “Diagnosis / Error Messages / Warnings“
Malfunctions listed below can probably be cleared onsite.
Symptom Possible cause Action
No display on the LC
display (option)
No supply voltage
Fuse defective
Check voltage supply.
Contact SICK Service.
Message Significance Possible cause Action
AO configuration Number of analog outputs
configured and number of optional
modules do not match.
AO not configured
Connection error
Module failure
Check configuration (
p. 73,
§4.2.6).
Contact SICK Service.
AI configuration Number of analog inputs configured
and number of optional modules do
not match.
AI not configured
Connection error
Module failure
Check configuration (
p. 75,
§4.2.7).
Contact SICK Service.