EasyManua.ls Logo

SICK FLOWSIC200 - Standard Commissioning Procedure

SICK FLOWSIC200
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning and parameterization
FLOWSIC200 · Operating Instructions · 8013271/1CJ9/V2-0/2022-01 · © SICK Engineering GmbH 67
The level 1 default password is included in the Annex.
Fig. 53 Entering the password
4.2 Standard commissioning procedure
This Section describes all the settings required for the device function.
To set/change the parameters, carry out the following procedure:
Connect the measuring system to the SOPAS ET program, scan the network and add the
required device file ( MCU, FLOWSIC200 M/FLOWSIC200 H/FLOWSIC200 H-M ) to the
current project (

p. 64, §4.1.4).
Enter the Level 1 password (

p. 64, §4.1.4) and open directory “Maintenance /
Maintenance / Operation”.
Activate checkbox “Maintenance” (MCU) or “Sensor maintenance” (sender/receiver
unit) and click “Set State”.
Fig. 54 Switching to Maintenance mode
Error message “Error Parameter” is output as long as the plant data (

p. 71,
§4.2.4) have not been entered completely.

Table of Contents

Related product manuals