EasyManua.ls Logo

SICK FLOWSIC200 - Output Flow Direction

SICK FLOWSIC200
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
78 FLOWSIC200 · Operating Instructions · 8013271/1CJ9/V 2-0/2022-01 · © SICK Engineering GmbH
Commissioning and parameterization
4.2.10 Output flow direction
To output the flow direction, the measuring range must be set in a negative and a positive
range in the “Configuration / I/O Configuration / Analog Output / Analog Outputs” directory.
The zero point then lies between the two end values. Falling below or exceeding the zero
point can additionally be signaled with the limit value relay (

p. 76, §4.2.8).
Example:
Measuring range: -20 ... +20 m/s
The Live zero value (= lower end value) corresponds to a physical value of -20 m/s, the
20 mA value (= upper end value) corresponds to a physical value of +20 m/s.
With Live zero = 4 mA, the zero point is at 12 mA.
Fig. 65 Directory “Configuration / I/O Configuration / Analog Output / Analog Outputs”
Optionally, only the amount of the measured value can be output. For this purpose, the
“Write absolute value” checkbox must be activated.
NOTICE:
The limit value relay must be configured when the absolute value is output,
otherwise no signaling of falling below or exceeding the zero point is possible.

Table of Contents

Related product manuals