4-69
1010GCNFM-3C
Section 4
4.11.6 THE TEST FACILITIES MENU
The Test Facilities menu provides commands for system analysis and recovery. The most useful for the
end-user are Makeup and Graph. Using these routines under the supervision of our technical support
staff will help us to provide technical analysis and solutions.
To start a Test Routine:
1. Use the <Up or Down Arrow> to scroll the cursor to the test routine you want to use.
2. Press the <Right Arrow> to access the test option list. Use the <Up or Down Arrow> to move the
asterisk to [Yes].
3. Press <ENT> to start the routine.
4. Press <Left Arrow> to end the routine.
Siemens Dual Path SITE1
Makeup No
Graph No
Tx Up No
Tx Dn No
Fixed ALC No
Tx Up Fixed ALC No
Tx Dn Fixed ALC No
Graph AutoZero No
Right arrow invokes an InProcess Makeup
Test Facilities
* For technical service diagnostics.
Makeup Commands the meter to re-acquire the receive signal.
Graph Enables an on-screen marker display.
Tx Up Forces upstream transmission only.*
Tx Dn Forces downstream transmission only.*
Fixed ALC Select “Yes” to command amplifier to hold a constant gain.*
TX Up Fixed ALC Fixed ALC while transmitting up only.*
TX Dn Fixed ALC Fixed ALC while transmitting down only.*
Graph AutoZero Waveform of ZeroMatic signal reference.
TEST FACILITIES COMMANDS