4-32
1010GCNFM-3C
Section 4
When the inital makeup of ZeroMatic is complete the screen will return to the Pick/Install Xdcr menu and
automatically highlight [Operation Adjust], which is the next menu cell.
Channel/Path Setup
Pipe Data
Application Data
Pick/Install Xdcr
Operation Adjust
Flow/Total Units
Data Span/Set/Cal
Stripchart Setup
Datalogger Setup
I/O Data Control
Diagnostic Data
Siemens Dual Path SITE 1
Dual Path Flow
Conforms operation to user Preferences
To disable the ZeroMatic function:
Select the [Pick/Install Xdcr] menu cell from the Dual Channel Flow menu.
Scroll down to the [Zero Flow Adjust] menu cell by pressing
NOTE: The highlighted [ZeroMatic] menu item is the only indication that ZeroMatic is function-
ing.
Invoke the ZeroMatic initial makeup procedure as previously described above.
While ZeroMatic initial makeup is running, press to abort the process and thereby disabling
the function.
The screen will return to the Dual Channel Flow menu and highlight the [Operation Adjust] menu
cell.
4.5 THE OPERATION ADJUST MENU
This menu becomes available after picking a meter type and measurement channel. We recommend
that you use it after the transducers are installed and operating to “fine-tune” the meter’s output charac-
teristics.
Each application presents different data display and output requirements due to unique pipe and liquid
conditions. Use the Operation Adjust menu to match meter operation to the site. You can set damping
controls for the primary flow rate output. You can define a Deadband, (usually a very low flow rate),
below which, the flow output will be forced to zero. You can also select the meter response to a continu-
ous Fault condition.