4-26
1010GCNFM-3C
Section 4
NOTE: The <ALT> and <MENU> keys must be pressed before the flowmeter scans through all
the drives, or the selection of the detection mode and the Force Transmit function must
be initiated again.
2. A typical menu screen will appear as shown below and indicate the current ALC (e.g., 50).
This ALC number indicates the current receive signal strength and can be used for further diagnostic
purposes.
Force Transmit Procedure
This diagnostic software routine allows the user to “force” a transmitting condition that can be use to
search for an amplitude level (ALC) when Detection Fault or Low Signal alarms are present. The routine
forces the flowmeter to generate constant transmit bursts while reporting current receive signal strength
for the user. To initiate the Force Transmit function, refer to the example shown below.
1. After [Install] is selected press <ENT>. While the flowmeter is going through the drives (see menu
screen below), press the <ALT> and <MENU> keys simultaneously.
Install Path 1
Transducer Model 1011G Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? Install
Zero Flow Adjust Channel Not Setup
Siemens Dual Path SITE1
Drive 11
Pick/Install Xdcr
Install Path 1
Transducer Model 1011G Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? No
Zero Flow Adjust Channel Not Setup
Siemens Dual Path SITE1
Key Install after mounting transducers
Pick/Install Xdcr
Install Path 1
Transducer Model 1011G Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? Install
Zero Flow Adjust Channel Not Setup
Siemens Dual Path SITE1
ForceN fx=8 m=7 ALC=50
Pick/Install Xdcr
Install Path 1
Transducer Model 1011G Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset
Number Index
Spacing Method
Ltn Value (in)
Install Completed? Install
Zero Flow Adjust Channel Not Setup
Siemens Dual Path SITE1
Pick/Install Xdcr
Detection Fault
Press [ENT]