4-25
1010GCNFM-3C
Section 4
After transducers are properly mounted, the flowmeter completes its Initial Makeup command and the
[Install Completed?] menu cell appears.
z To access the [Install Completed?] option list, press the <Right Arrow>.
z Press <Down Arrow> and scroll to the [Install] menu cell and then press <ENT> (unless other-
wise directed to do so by the Technical Service Department).
z Press [ENT]. The [Install Completed?] menu cell will indicate [Yes] after the FUH1010GCN is
successfully installed.
NOTE: If the transducers have been installed successfully but the Estimated Vs (sonic velocity)
has been changed, the Pick/Install Xdcr menu cell [Install Completed?] option list will
also display the [Transfer Install] function selection. The Transfer Install function allows
the transducers to be optimally positioned for a different fluid, without the need for a
new Initial Makeup procedure (see the Water Calibration addendum in this manual for
details).
To complete the Install process disregard the following paragraphs explaining the Force
Transmit and Force Frequency diagnostic software routines. Proceed to the Install pro-
cess completion steps immediately following the Force Transmit procedures.
Install Path 1
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? Yes
Zero Flow Adjust Channel Not Setup
Siemens Dual Path SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Install Path 1
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? No
Zero Flow Adjust Channel Not Setup
Siemens Dual Path SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr