B-9
1010GCNFM-3C
Appendix
To Clear All Dynamic Memory Using the RS-232 Interface:
CAUTION: Before proceeding further , it is essential to understand that this function eliminates
all data stored in RAM. This means that all saved site setups, including the site data of
a flow-calibrated site will be erased! Also, the entire Datalogger file plus any custom
factory or user-created pipe or transducer tables will be eliminated. The impact of
this is such that we strongly recommend that you consult our technical service de-
partment before continuing with this procedure. If you choose to continue, be aware
that you will have to create a new site setup, re-enter all site specific parameters
including pipe or transducer tables, plus all desired Meter Facilities entries.
1. Turn off power (if it is currently on).
2. Turn the power on. As soon as you apply power, type the @ character three times.
The prompt: [Clr Active Memory?] appears at the top of the screen. Press the <Down Arrow>.
Note that the prompt switches to [Clr Saved Data? No] .
3. Press the <Right Arrow> and then the <Down Arrow> to switch the option list to:
[Clr Saved Data? Yes].
4. Press <ENT> to clear all Saved Site Data, Datalogger Data, user created Pipe Data and Transducer
Data.
5. To restore operation, press <MENU> to access the Installation Menu. Create a new site setup or recall
a stored site setup. Reselect any Meter Facilities items (e.g., RS-232 setup parameters).
RESET PROCEDURE FOR BLIND SYSTEMS
SITRANS F 1010 Blind systems allow you to perform a system reset via the RS-232 interface. The follow-
ing instructions require the flowmeter to be connected serially to a PC.
NOTE: Custom RS-232 settings for baud rate, parity and data bits may not be preserved. There-
fore, be prepared to set your communications program back to the default (9600, Odd,
7) settings.
To Clear Active Memory Using the RS-232 Interface
1. Turn off power (if it is currently on). Turn power on. As soon as you apply power, immediately type
the @ character three times. The prompt: [Clr Active Memory? No] appears at the top of the
screen.
2. Press the <Right Arrow> and then the <Down Arrow> to switch the option list to:
[Clr Active Memory? Yes] Press <ENT> to clear all Active Site Data (but not saved site setups).
3. To restore operation, press <MENU> to access the Installation Menu. Create a new site setup or recall
a stored site setup. Re-select any Meter Facilities items (e.g., RS-232 setup parameters).