Note
Selecting the [Operation Adjust] menu cell [Zero/Fault Set] allows the option of indicating
Zero Flow rather than a Fault.
Selecting Liquid Composition
The width of the Spectra wave shape is an indication of the depth of penetration of sonic energy
into the flow stream. The % Deviation is a measure of the width of the Spectra and is
representative of the penetration of the sonic beam into the liquid.
Select [Slurry] in the [Application Data] menu for better flow rate indication accuracy when the
% Deviation is less than 25 with typical flow conditions.
For % Deviations above 35, select [Liquid]. For applications that continuously have %
Deviations between 25 and 35, proper selection of either [Liquid] or [Slurry] can be confirmed
by comparison to a known flow reference.
Other Menu Entries
All other flowmeter menu entries will operate and be used in the same manner as the transit
time flow measurement mode.
Functions
7.11 Reflexor
FUP1010 IP67 Portable Flowmeter
108
Operating Instructions, 02/2010, A5E02951522A Revision 01