Pin# Signal Function Description
D N/C N/C No Connection
E R2B Logic Out 2 Logic Level State (HIGH=5 to 3 Vdc; LOW= 1 to 0 Vdc)
F N/C N/C No Connection
G R3B Logic Out 3 Logic Level State (HIGH=5 to 3 Vdc; LOW= 1 to 0 Vdc)
H N/C N/C No Connection
J R4B Logic Out 4 Logic Level State (HIGH=5 to 3 Vdc; LOW= 1 to 0 Vdc)
K PGEN1 Frequency Output 1 Assignable Logic Level Pulse Train (5V TTL CMOS Logic)
L PGEN2 Frequency Output 2 Assignable Logic Level Pulse Train (5V TTL CMOS Logic)
M Vo1 Voltage Output 1 Assignable (0 to 10 Vdc) [Min. load=5k ohms]
N Vo2 Voltage Output 2 Assignable (0 to 10 Vdc) [Min. load=5k ohms]
P Io1 Current Output 1 Assignable 4 to 20mA output
R Io2 Current Output 2 Assignable 4 to 20mA output
S Io PWR External Current Loop User supplied 18 to 30 Vdc @ 25 mA min.
T GND Signal Ground Return For Signals
U CNTRL1 Logic In 1 TTL Level State [HIGH=5 to 3 Vdc; LOW=1 to 0 Vdc]
V CNTRL2 Logic In 2 TTL Level State [HIGH=5 to 3 Vdc; LOW=1 to 0 Vdc]
W CNTRL3 Logic In 3 TTL Level State [HIGH=5 to 3 Vdc; LOW=1 to 0 Vdc]
X CNTRL4 Logic In 4 TTL Level State [HIGH=5 to 3 Vdc; LOW=1 to 0 Vdc]
Y Iin1 Current Input 1 External sensor input (4-20 mA) [Load = 250 ohms]
Z Iin2 Current Input 2 External sensor input (4-20 mA) [Load = 250 ohms]
a Vin1 Voltage Input 1 External sensor input (0 to 10 Vdc) [Load = 100k ohms]
b Vin2 Voltage Input 2 External sensor input (0 to 10 Vdc) [Load = 100k ohms]
c GND Signal Ground Return For Signals
A.4 Site Setup For SITRANS F
How to use the Windows HyperTerminal Program
Windows provides a communication program called HyperTerminal, which is ideal for
interfacing your computer with the flowmeter. The following typical example explains how to
set up HyperTerminal.
Note
Depending upon the Windows applications being used this setup procedure may vary.
1. From the Windows desktop, left-click on the [START] button.
2. Holding down the left mouse button, move the highlight up to [Programs], then across to
[Accessories] then [Communications]. Slide the highlight down to [HyperTerminal], then
release the left mouse button.
3. Within the HyperTerminal window, move the mouse pointer down to [Hyperterm.exe] and
then double-click the left mouse button.
Appendix
A.4 Site Setup For SITRANS F
FUP1010 IP67 Portable Flowmeter
138
Operating Instructions, 02/2010, A5E02951522A Revision 01