To activate the Test Facilities Graph Screen:
1. In the main menu, scroll to the [Diagnostic Data] menu and select [Test Facilities].
2. Scroll
down to [Graph], press the <Right Arrow> and highlight [Yes]. Press <ENT> to select.
3. The Test Facilities Graphic Screen will appear on the meter display as shown below.
Damping Factors Min Damping Factor (Hot Key 1)
Digitized Receive Signal TN Marker
Crossover Marker High Baseline Noise
Figure 10-2 Setting Digital Damping Factor
Setting the Digital Damping Factor to a value HIGHER than the default value of 4 may be
necessary in cases where the signal-to-noise ratio (SN) is found to be unacceptably low
(<15:1), but only if the noise is determined to be asynchronous (i.e., not associated with the
transmit or flowmeter timing circuitry) as shown in the signal example above, where the
baseline noise has a higher frequency than the true liquid signal.
The following application conditions may require a higher Digital Damping Factor:
Close proximity to pressure control valves which may generate in-band acoustic noise
High un-dissolved gas solids content in liquid.
High electronic noise from variable frequency drives or other external equipment.
Troubleshooting
10.3 Test Facilities Graph Screen
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
123

Table of Contents

Other manuals for Siemens SITRANS FUP1010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SITRANS FUP1010 and is the answer not in the manual?

Siemens SITRANS FUP1010 Specifications

General IconGeneral
BrandSiemens
ModelSITRANS FUP1010
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals