Sensor cable connection Customer's Pipe
1011HPS Series Sensor Sensor Index Screw
Spring Clip Sensor Index Screws-Storage Location
Spacer Bar 1011PPS Series Sensor
Figure 5-8 Reflect Mount with EZ-Clamp and Spacer Bar
Installation Procedure
1. After receiving the
Number Index from the Installation Menu, prepare the pipe surface area
where the sensors will be mounted. Check to ensure that you have a matched set of
sensors. They both should have the same S/N number but marked with either an "A" or "B"
(e.g., 100A and 100B).
2. Degrease the surface and remove any grit, corrosion, rust, loose paint, etc.
3. Assemble the sensors to the spacer bar, with the cable connectors facing away from each
other as shown above. The spacer bar is attached to a sensor using a sensor index screw.
Connecting
5.2 Sensor Wiring
FUP1010 IP67 Portable Flowmeter
38
Operating Instructions, 02/2010, A5E02951522A Revision 01