ZeroMatic (optional function)
Note
ZeroMatic is used
in the Reflect Mode only. Invoking ZeroMatic clears any existing fixed zero
offset.
Use this menu cell to select the ZeroMatic option. If conditions permit the use of the Auto Zero
function then the ZeroMatic option can be used as well.
To select and enable the ZeroMatic option:
1. In the Install Xdcr menu, press <Up Arrow> to scroll to the [Zero Flow Adjust] menu cell.
2. To access the [Zero Flow Adjust] option list press <Right Arrow>.
Note
If ZeroMatic is
not running, the [Actual Zero] menu item will be displayed next to the [Zero
Flow Adjust] menu cell.
3. Select the [ZeroMatic] menu cell by pressing <Up/Down Arrow> then press <ENT>.
When the Initial
Makeup of ZeroMatic is complete the screen will return to the Pick/Install Xdcr
menu and automatically highlight [Operation Adjust], which is the next menu cell.
To disable the ZeroMatic function:
1. Select the [Install Xdcr] menu cell from the Dual Path Flow menu.
2. Scroll down to the [Zero Flow Adjust] menu cell by pressing <Up/Down Arrow>.
Note
The highlighted [ZeroMatic]
menu item is the only indication that ZeroMatic is functioning.
3. Invoke the ZeroMatic initial makeup procedure as previously described above.
4. While ZeroMatic initial
makeup is running, press <Left Arrow> to abort the process thereby
disabling the function.
5. The screen will return to the Dual Path Flow menu and highlight the [Operation Adjust]
menu cell.
Connecting
5.3 Zero Flow Adjust Menu
FUP1010 IP67 Portable Flowmeter
Operating Instructions, 02/2010, A5E02951522A Revision 01
63