1250 Frequency Response Analyzer The RS423 and GPIB Ports. Remote Control
AMK/1250/4 13-15
Facility Parameter Code Arg Meaning Max Min Init
GPIB OUTPUT OS I 0=comma 1 0 comma
FORMAT SEPARATOR 1=terminator
GPIB OUTPUT OT I 0=crlf 3 0 crlf
FORMAT TERMINATOR 1=crlf + EOI
2=cr
3=cr + EOI
RS423/GPIB RR I 0=on 1 0 on
REFORMAT ENABLE 1=off
RS423/GPIB RH I 0=on 1 0 on
HEADING ENABLE 1=off
13.7.17 FUNCTION COMMANDS
Facility Parameter Code Arg Meaning Max Min Init
ANGLE RANGE FN 1, I 0= - to +
1=0 to 2
2= -2 to 0
3=0 to 2 (Clk)
3 0 - to +
ANGLE UNITS FN 2, I 0=deg
1=rad
1 0 deg
OMEGA SCALE FN 3, I 0=off
1=x jω
2= / jω
3=x ( jω )
2
4= / ( jω )
2
4 0 off
VECTOR SCALE FN 4, I 0=off
1=a, jb
2=r, θ
3=last result
3 0 off
LIMIT CHECK FN 5, I o=off
1=a
2=b
3=r
4=θ
5=log r
5 0 off
SCALE FACTOR
a
AF F 1E18 -1E18 ne
SCALE FACTOR
b
BF F 1E18 -1E18 ne
SCALE FACTOR
r
RF F 1E18 -1E18 ne
SCALE FACTOR
θ
TF F +180 -180 ne
UPPER LIMIT
a
UA F 1E18 -1E18 ne
UPPER LIMIT
b
UB F 1E18 -1E18 ne
UPPER LIMIT
r
UR F 1E18 -1E18 ne
UPPER LIMIT
θ
UT F +180 -180 ne
UPPER LIMIT log r UG F +380 -380 ne
LOWER LIMIT
a
LA F 1E18 -1E18 ne
LOWER LIMIT
b
LB F 1E18 -1E18 ne
LOWER LIMIT
r
LR F 1E18 -1E18 ne
LOWER LIMIT
θ
LT F +180 -180 ne
LOWER LIMIT log r LG F +380 -380 ne
13.7.18 MISCELLANEOUS COMMANDS
Facility Parameter Code Arg Meaning Max Min Init
LOCAL LL none (Allowed from
RS423 only)
none