Additional Displayed Information 1250 Frequency Response Analyzer
11-8 AMK/1250/3
GPIB (cont.)
Soft Keys Page Display
2 GPIB SER. POLL PAR. POLL CONTROL
8 digits, consisting A number local/
of a set of “flags” from 0 to 8 remote/
and a “service” 0 = disabled loc + llo/
bit. rem + llo
followed by sense:
T = True
F = False
Note: SER = Serial PAR = Parallel llo = Local lock-out
For further explanation of these messages, see the GPIB section,
Chapter 13, Section 7.
3 GPIB I/P TERM O/P TERM O/P SEP
lf/cr/s-colon/EOI crlf/crlf+EOI/ comma/term
cr/cr+EOI
Notes: TERM = Terminator, SEP = Separator
cr = carriage return, lf = line feed, EOI = End or Identify.
Where "term" appears under 0/P SEP, it means that the Output
Separator is the same as the Output Terminator. Further information
appears in the GPIB section, Chapter 13.
11.1.5 PRINTED RECORD
If a printed record is required of the STATUS information, enable the relevant output
device as described under DATA OUTPUT. Any state other than off will cause the data
to be printed; the pass/fail conditions do not apply here. Note that STATUS information
is not available via the GPIB.
Press STATUS, then the relevant soft key. It is only necessary to use SELECT to
change from [MAIN] to [OTHER] STATUS; the pages will be "scrolled" automatically to
the output device when the soft key is pressed. Note, however, that only the first page
for each soft key will appear in the Display.