Tidal volume Vt = 10 ml
Respiratory rate f = 40 1/m
Tinsp = 0.4 s
TRamp = 0.4 s
Basic flow at 40% oxygen = 3 l/min
The minute volume is calculated by multiplying the tidal volume with the
respiratory rate:
At very high temperatures the basic flow may rise, but due to the blower
module the EVE
TR
can provide ventilation without any added oxygen at
any time.
5.1.3 Changing the oxygen cylinder
Before changing the pressurized gas cylinder, refer to the operating manual of the
pressure regulator.
Sample calculation:
Neo patient supplied
with 95% oxygen