11 Care and maintenance
226
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
11.8 Treatment table
System components
Recommended
treatment intervals
Automated
thermal
treatment
Manual
treatment
Sterilization
Case, power cord,
connection tubes
After every patient/
weekly
No
Surfaces
No
Touchscreen
After every patient/
weekly
No
Surfaces
No
Docking station
After every patient/
weekly
No
Surfaces
No
Ambulance bracket
If contaminated
No
Surfaces
No
Helicopter bracket
If contaminated
No
Surfaces
No
Carry system
If contaminated
No
Surfaces
No
Accessories bags
(carry system)
If contaminated
No
Clean manually
using a brush
No
Adult test lung
After every patient/
weekly
Observe the manufacturer’s operating
manual!
Test lung Neo with
tube adapter
After every patient/
weekly
Yes
Yes
Yes
Reusable
expiration valve
After every patient/
weekly
Yes
Yes
Yes
Reusable patient
tube system
After every patient/
weekly
Yes
Yes
Yes
Type B
pneumotachograph
After every patient/
weekly
Yes
Yes
Yes