EasyManua.ls Logo

Stephan EVETR - Non-Invasive DUOPAP (Nduopap)

Stephan EVETR
254 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Ventilation modes
162
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Non-invasive DUOPAP (nDUOPAP) 7.3.2.10
nDUOPAP corresponds to the behavior of the DUOPAP. It is only
available in non-invasive Neo mode. The device is operated without a
flow sensor (PNT) in this mode. The pressure is measured proximally via
the pressure measurement adapter set.
Ventilation mode
Adjustable parameters
nDUOPAP
O
2
Inspiration time (Tinsp) and breathing frequency
(f)
(the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Positive end expiratory pressure (PEEP)
Ramp time
Trigger
Flow limitation (in NEO mode only)
Tab. 35: nDUOPAP
Ventilation mode
Adjustable parameters
nDUOPAP
with PSV
O
2
Upper pressure level (Phigh)
Inspiration time (Tinsp) and breathing frequency
(f) (the parameters for breathing cycling can be
configured, see chapter 4.4.2)
Positive end expiratory pressure (PEEP)
Ramp time
Pressure support (Psupp)
(relative set value with reference to PEEP)
Trigger
Tab. 36: nDUOPAP with PSV