10 Troubleshooting
208
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
10.2 Selftest error
Error during
power supply test
HW
Battery malfunction
Battery empty
Check battery level, charge
battery if necessary
If the error cannot be
corrected, contact
FRITZ STEPHAN GMBH
customer service, providing
the error code displayed
Error during valve
test
HW
Valve malfunction
Contact
FRITZ STEPHAN GMBH
customer service, providing
the error code displayed
Error during sensor
test
HW
Sensor system malfunction
Check secure connection of
the patient tube system
If the error cannot be
corrected, contact
FRITZ STEPHAN GMBH
customer service, providing
the error code displayed
Error during
turbine test
HW
Turbine malfunction
Check secure connection of
the patient tube system
Check system for leakage
Check sealing plug
on Y piece
If the error cannot be
corrected, contact
FRITZ STEPHAN GMBH
customer service, providing
the error code displayed
Tab. 45: Selftest error