7 Ventilation modes
142
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
7.2 Trigger functionality
7.2.1 Flow trigger
The flow sensor enables the EVE
TR
to measure the patient’s inspiratory
flow. If the inspiratory flow exceeds the value set by the user, a
mandatory breath is triggered. This "trigger threshold" is set in the
»Trigger« field as flow in l/min.
7.2.2 Internal flow trigger
The internal flow trigger is active for all non-invasive neonatal
ventilation forms (optional). The "trigger threshold" is set in the
»Trigger« field as flow in l/min.