Do not use any disinfectants based on amines or their derivatives! These agents can
damage the plastics and silicone in the device and cause malfunctions.
Do not use any disinfectants containing oxygen, chlorine or halogen donors or
phenols and their derivatives! These agents can damage the material and cause
malfunctions.
Do not use any highly alkaline cleaning agents (pH > 10.9)! These agents can damage
the material and cause malfunctions.
Do not use any rinse aids or other drying agents!
11.3 Automated cleaning and disinfection
For automated cleaning, configure the automated cleaning and
disinfection equipment according to the manufacturer’s operating
instructions.
Use only automated cleaning and disinfection equipment that complies with
DIN EN ISO 15883-1!
The parts to be cleaned must be disassembled according to the instructions
in this operating manual before treatment. Load the automated cleaning
and disinfection equipment in such a way that the inner and outer
surfaces of the parts can be reached by the cleaning agent and no spots
are missed.