6 Operation
132
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
6.7.1 Default ventilation parameters
The ventilator offers default ventilation parameters for all ventilation
modes for the three patient types. These parameters are activated after
selecting the type. They can be adapted to the corresponding patient in
the Mode Settings menu.
Parameter
Preterm infants and
newborns (optional)
Child
Adult
Apnea
4 s
4 s
4 s
ETS
25%
25%
25%
O
2
21%
21%
21%
f
40
20
12
I:E
1:2.7
1:1.7
1:1.7
PEEP
5 mbar
5 mbar
5 mbar
Phigh
18 mbar
12 mbar
12 mbar
Pinsp
18 mbar
12 mbar
12 mbar
Psupp
1 mbar
10 mbar
10 mbar
Ramp time
0.1 s
0.2 s
0.2 s
VT
5 ml
200 ml
500 ml
Trigger
0.5
3.0
4.0
Flow limit
9 l/min
--
--
Tab. 14: Default ventilation parameters