11 Care and maintenance
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
217
All disinfectant residue must be completely removed afterwards.
Suitable cleaning agents with regard to material compatibility are
aldehyde-free quick disinfecting wipes, such as Bacillol
®
wipes or tissues
made by Paul Hartmann AG, Heidenheim or Mikrozid
®
PAA wipes
made by Schülke, Norderstedt.
11.6 Sterilization
Components labeled in this operating manual as suitable for sterilization
can be sterilized with hot steam at temperatures of up to 134° C.
The hot steam sterilizer must be loaded according to the manufacturer’s
instructions for use and the corresponding program started.
Temperature
Hold/dwell time
134° C
3 18 min.
Tab. 47: Standard procedure for steam sterilization*
*All standard steam sterilization procedures can be used.