EasyManua.ls Logo

Stephan EVETR - Therapy; High Flow O Therapy

Stephan EVETR
254 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Ventilation modes
166
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
7.3.4 O
2
therapy
O
2
therapy is not a ventilation mode in the strict sense. Its purpose is
rather a dosed enrichment of the breathing air with oxygen in case of
breathing problems and to increase the O
2
level in the arterial blood. The
O
2
concentration and flow can be set on the ventilator. Monitoring by the
EVE
TR
requires the connection of an (optional) SpO
2
sensor. The oxygen
is delivered by a nasal tube, nasal cannula, or oxygen mask.
7.3.5 High flow O
2
therapy
High flow O
2
therapy is available only in Neo mode and is actually not a
form of ventilation; instead it serves as ventilatory support. Oxygen is
supplied using "high flow nasal cannula". O
2
concentration, flow, and
operating pressure limit (PAW alarm limit) can be set on the ventilator
The respirator monitors this only if an SpO
2
sensor (optional) is
connected.
Ventilation mode
Adjustable parameters
High-Flow
FiO
2
Flow
Pressure limit (PAW alarm limit)
Tab. 41: High-Flow

Table of Contents

Related product manuals