Use only approved patient filters. Otherwise, the patient may be at risk. Always
follow the manufacturer's operating manual!
5.7 Installing the SpO
2
sensor
The SpO
2
sensor for measuring the Masimo-SET Rainbow® parameters
is connected to the SpO
2
input on the right side of the EVE
TR
device
(see chapter 3.4).
5.8 Aerosol nebulization
Due to tolerances in the nebulizer flow, the minute and breathing volume displayed
during medication nebulization may differ significantly from the actual values. As a
result, FRITZ STEPHAN GMBH recommends using a pressure-controlled ventilation
mode for nebulization, especially in case of children.
Nebulization and humidification may increase the resistance of patient filters. The
user must regularly inspect the patient filter for increased resistance or blockage.
Select the medication dose carefully. If the medication dose is too high, it may
negatively affect the ambient air.
The accuracy of the ventilator may be impaired by the gas added by the nebulizer.