6.1.2 Test list
The operating manual is part of the
device and must always be kept with
the device.
Visual check of color-coding of gas
types
ISO color code
O
2
(oxygen) white
Unique mechanical features of the
angled connector and gas connections
O
2
pressurized gas cylinders
Leak tightness
Residual pressure
Visual check of the mains connection
Undamaged
Mains switch »On/Off« to "On"
»On« button illuminated
Patient connections
(All patient components
must be connected properly)
Check the patient tube system
Check the SpO
2
sensor (optional)
Check the CO
2
sensor (optional)
Check the flow sensor (PNT)
Check the expiration valve
Self-filling, present, within reach
Selftest completed successfully
Disconnect the device from the
external energy supply
Battery sufficiently charged
Check if the acoustic alarm is
triggered during the selftest
Acoustic alarm is triggered
during the selftest
Disconnect the device from the
external energy supply
"Powersupply?" notification
appears.
Disconnect the patient tube system
from the ventilator
"Disconnection!!!" HP alarm
appears
Close O
2
pressure regulator,
set O
2
level to 30%.
After 30 s, "O2 low!!" MP alarm
appears
Connect test lung, set PAW UL to
minimum, trigger peak pressure
Airway pressure high!!!
HP alarm appears.
Tab. 13: Test list