10 Troubleshooting
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
207
Int. power supply
failure
(Alarm delay: 2 s)
HP
TA
Internal power
supply malfunction
Contact
FRITZ STEPHAN GMBH
customer service
Occlusion
HP
TA
Airway blockage
Airway tube kinked
ETT kinked
The EVE
TR
automatically reduces the
pressure to PEEP level
and then opens the
expiration valve so that
the patient can breathe
freely.
Check the airway tube
Check the ETT
PEEP low
HP
PA
PEEP measured
value is below lower
PEEP alarm limit
Check inspiration and
expiration times
Check ventilation tube
system connections for
possible leakage
Check expiration valve
for proper connection
Check mask and tube for
proper connection
P-Meas.: Check VBS
HP
TA
Pressure
measurement failure
Check patient tube
system
Tab. 44: Troubleshooting