8 CO2 measurement (optional)
The ISA™ CO
2
analyzer may not be used with flammable anesthetics.
Make sure the sample collection tube is routed properly to prevent the risk of patient
entanglement or strangulation.
Do not use the sample collection tube configuration for adults and children with
infants because this may add 7 ml of dead space to the patient system circuit.
Do not use the sample collection tube configuration for infants with adults because
this may cause excessive flow resistance.
The Nomoline sample collection tube and its connections are non-sterile. To prevent
damage, do not autoclave any part of the sample collection tube.
The sample collection tube must be replaced if the tube input connector starts flashing
red or a Nomoline locking message appears on the EVE
TR
display.
The ISA™ CO
2
analyzer is not designed for use in MRI environments. During MRI
scans, the analyzer must remain outside the MRI room.