1 General information
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
25
nPC-ACV
Non-invasive pressure-controlled assist/control
ventilation
nPC-ACV+
Non-invasive pressure-controlled assist/control
ventilation with expiration trigger
nDUOPAP
Non-invasive ventilation support at two different
pressure levels
nCPAP
Non-invasive spontaneous breathing with continuous
positive airway pressure
Parameter
Setting range
Resolution
Pre-oxygenation (time)
10 180 s
1
Pre-oxygenation (concentration)
21 100%
1
Insp. Hold
Max. 15 s
1
Aerosol
5 30 min
5
Parameter
Meaning
Setting range
Resolution
PEEP
Positive end expiratory
pressure
0 25 mbar
1
Pinsp*
Inspiratory pressure
6 55 mbar
1
Psupp
Support pressure
1 55 mbar
Phigh*
Inspiratory pressure
under DUOPAP
6 55 mbar
1
VT
Tidal volume
(PC mode / PRVC)
2 150 ml
1
150 500 ml
5
500 1000 ml
10
1000 2000 ml
50
VT
Tidal volume
(volume-controlled
ventilation)
100 150 ml
1
150 500 ml
5
500 1000 ml
10
1000 2000 ml
50
Flow limit
Flow limitation
5 -30 / Off l/min
1
Tinsp
Inspiration time
0.2 30 s
0.1
Texp
Expiration time
0.2 30 s
0.1