1 General information
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
27
Measured value
Unit
Resolution
Display range
Time (accuracy
1
: t` = 5%)
Tinsp
s
0.1
0 60
Texp
s
0.1
0 60
ftotal
1/min
1
0 300
fspon
1/min
1
0 300
I:E
I:E
0.1
1:150 150:1
O
2
(calculation accuracy
1
: 2.5 Vol.% + 2.5% of displayed value)
O
2
%
1
21 100
Rise time of O
2
level from 21% to
90%
58 s (VT = 30 ml)
37 s (VT = 150 ml)
22 s (VT = 500 ml)
EtCO
2
(see manufacturer data sheet for sensor accuracy)
EtCO
2
Vol%
0.1
0 90
mmHg
1
0 12
kPa
0.1
0 999
Lung dynamics
R
mbar/l/s
1
0 1000
C
ml/mbar
0.1
0 650
Diagnostic
P0.1
mbar
1
0 -25
Time constant
s
0.1
0 20
RSB
1/min*L
1
0 9999
PTP
mbar x s
0.1
0 999
MASIMO® parameters
(see manufacturer data sheet for accuracy of MASIMO
®
sensors)
Pulse
bpm
1
0 239
(above "---")
PVI
%
1
0 100%
PI
%
0.01
0.02 0.99
0.1
1 9.9%
1
10 20%
SpM et
%
0.1
0 100
SpCO
%
0.1
0 100
SpOC
ml/dl
0.1
0 35