7 Maintaining the product
0970 4020 en 01 135
7. Enter the Measuring cycle and determine its Unit.
The measuring cycle determines in which intervals a
new reading is saved in the Saveris base.
Possible settings for the unit:
sec (second)
min (minute)
h (hour).
The smallest transfer rate for radio probes is one
minute.
8. Determine Alarm delay.
If you enter the value "5" for example, an alarm is not
triggered by the Saveris base until the fifth time that the
limit value is exceeded.
9. Determine in which cases a system alarm should be
triggered.
10. Click on [Continue >].
- If a router is connected at the Saveris base the configuration
of the connection type for the probes is shown.
If no router is connected continue with step 15.