5 First steps
0970 4020 en 01 45
8. Enter Measuring cycle and determine its Unit.
The measuring cycle determines in which intervals a
new reading is saved in the Saveris base.
Possible settings for the unit:
sec (second)
min (minute)
h (hour).
The smallest transfer rate for radio probes is one
minute.
Later the settings can be changed for every probe
separately in the software.
9. Determine Alarm delay.
If you enter the value "5" for example, an alarm is not
triggered by the Saveris base until the fifth time that the
limit value is exceeded.
Later the settings can be changed for every probe
separately in the software.
10. Determine the issue options for the limit value alarms.
Possible options:
SMS (optional with Saveris base with GSM module)
E-mail
Relay of the Saveris base.
The system alarms are above all used for pointing out
irregularities in the measurement system. If no reaction
follows the system alarms, uninterrupted data recording
is not guaranteed.
If a system alarm is issued via
SMS or E-mail time
delays may occur.
The notification function via
SMS is only available if the
Saveris base is equipped with a GSM module.
Later the settings can be changed for every probe
separately in the software.

Table of Contents

Other manuals for TESTO Saveris

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO Saveris and is the answer not in the manual?

TESTO Saveris Specifications

General IconGeneral
Humidity Range0 to 100 %RH
Measurement ParametersTemperature, Humidity
Data TransmissionWireless
Alarm FunctionYes
Resolution0.1 °C, 0.1 %RH
Radio Frequency868 MHz, 915 MHz
Data StorageInternal memory, Cloud
Power SupplyBattery

Related product manuals