Preventive Maintenance Guidelines
Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution Administrator Guide v 2.2 125
Ratings: E = Excellent, G = Good, F = Fair, P = Poor
CAUTION: Do not allow liquid to penetrate the openings on CO
2
sensors.
22.1.3.3 General Precautions • The machine running Smart-Vue Server software runs continuously,
24/7/365, and should be connected to an Uninterruptible Power Supply
(UPS) that protects against power surges and provides temporary power
to back up the machine and components.
• All Smart-Vue products that use a power adapter should also be plugged
into a UPS.
• Weekly manual system testing (e.g., disconnect a sensor from its
Smart-Vue module, verify that the alarm system is working) should be
performed as defined in your Standard Operating Procedure (SOP).
• Recommended maintenance and calibration procedures should be
followed.
• If you are storing cold products, Thermo Scientific recommends use of a
back-up cooling system (e.g., CO
2
or LN
2
) to maintain freezer chamber
temperature below the critical level should a power failure occur.
Contact your local sales representative for more information.
Weak
Acids
Strong
Acids
Weak
Alkalis
Strong
Alkalis
Organic
Solvents
Alcohols
Hydro
Carbons
Fuels
Gamma
Radiation
UV
Radiation
Module casing: Polycarbonate/ABS Alloy
PC/ABS
E G G F PPPPG F
Seal/button: Saturated Styrenic Block
Copolymer Thermoplastic Elastomer SEBS
E G E G P FPPP F
Transparent Window Acrylic/Polycarbonate
Alloy PC/PMMA
G G G G P FPF F F
Temperature
sensor plug
CO
2
detector
opening
Power/communicati
on cable plug