Power-over- Ethernet (POE) TCP/IP Receiver
72 Administrator Guide v 2.2 Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution
CAUTION: We recommend using a TCP port number such as 19001 for
your first network receiver, 19002 for the second, etc. It is best for each
network receiver to use a different Local Port, however the system will
function if the Local Port is set to the same value on all modules.
3. In the Host zone (Figure 40), enter the following information:
Address IP address or name of the DNS server hosting the
Smart-Vue Server application.
Port TCP connection port on the server computer.
Protocol TCP
TCP Keep Alive 45000
Reconnect Timer 5000: in case the network receiver is disconnected from
the server, it will try to reconnect every five seconds
Flush Serial Enabled: turns on “tunnel” operation, whereby any
data arriving on the serial port is immediately
transmitted.
4. Important: Click on Submit to save your changes.
5. You may close the browser window while the PoE network receiver
restarts.
13.10 Checking DNS
Configuration (Optional)
If you are using your PoE network receiver with a Domain Name Server
(DNS), make sure the DNS configuration information is correct:

Table of Contents

Other manuals for Thermo Scientific Smart-Vue

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Smart-Vue and is the answer not in the manual?

Thermo Scientific Smart-Vue Specifications

General IconGeneral
BrandThermo Scientific
ModelSmart-Vue
CategoryControl Unit
LanguageEnglish