Power-over- Ethernet (POE) TCP/IP Receiver
Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution Administrator Guide v 2.2 71
13.9 Configuring the
Communication Mode
(“Outgoing” or
Auto-Connect
“Inbound”)
As described in the Smart-Vue Client software User Manual, a TCP/IP
receiver may be configured to receive connections automatically from
Smart-Vue Server, or to initiate connections with Smart-Vue Server. Follow
these instructions to prepare the receiver for use with Smart-Vue Server.
1. In the left-hand menu bar, click on Tunnel Tunnel 1 Connect
Mode:
Figure 41. TCP connection mode configuration
In this screen, you will need to adjust settings under both Accept Mode and
Connect Mode.
2. Depending on your configuration, enter the following information in
the above window:
Smart-Vue Server initiates
connections
Label in Smart-Vue Client
F11
F11 Create a
receiver: “Outgoing
connection”
Receiver initiates connections
Label in Smart-Vue Client
F11
F11 Create a
receiver: “Incoming
connection (Auto-connect)”
Accept Mode Always Disable
Connect mode Disable Always
Local Port Define the TCP port of your choice, such as 19001, 19002, etc.