• There may be a network communication problem, possibly related to a
firewall or TCP port conflict. Check with your IT department.
When I try to add the receiver in Smart-Vue Client, why do I get a
message saying that the “port number is already in use”?
Make sure you do not have any other receivers using the specified TCP port.
By default, the first port to specify in the TCP port for Smart-Vue Server
field in Smart-Vue Client is 11001. You must increment this number so
each receiver in your network has a unique TCP port.
12.14 Specifications • ISM (Industrial Scientific Medical) band with 4 frequencies: US/CAN
915 MHz; Europe 868 MHz; APAC 434 MHz; India and other
countries 867 MHz
• Channel width: 50 kHz
• Frequency deflection: 16 KHz
• Wireless transmission speed: 9600 Baud in NRZ mode
• Modulation type: GFSK
• Driven receiver sensitivity for BER = 1%: -107dbm to -110 dBm
• Driven transmission: 8 dBm to 10 dBm
• Power output: 25 mW or 25 mW/500 mW with automatic switching
(depending on model) for 868, 867, and 915 MHz; 10 mW for 434
MHz
• Power adaptor with 1.85 meter (6 ft) cable length and four different
international plugs, 50/60 Hz, 100-240 V AC
• Ethernet 10/100 Mbit (RJ-45)
• Casing: ABS plastic with IP54 protection for indoor use
• Operating conditions: 0°C to +50°C (+32° to +122°F); 0 to 90% RH,
non- condensing
• Size: 190 x 52 x 30 mm (7.4 x 2 x 1.2 in.)
• Weight: 165 g (5.8 oz.)