Troubleshooting
140 Administrator Guide v 2.2 Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution
23 Troubleshooting
23.1 Technical Alarms Complete list
Datalogging stopped (different from configuration) [-991]
Check to see if datalogging is enabled on the wireless module.
Run an on-demand read to check wireless communication and
sensor status.
Transfer settings to the module once again.
No communication with telephone modem
Network unavailable (no communication with IP dry contact)
Wireless connection with siren temporarily interrupted [-994]
Indicates an error with an alert device.
For sirens: make sure that the device is properly connected to its
power supply and that it contains a sufficiently charged battery;
check the wireless connection between the receiver and wireless siren
(in Smart-Vue Client, under Add alert device – F8).
For a telephone modem: make sure that the modem is properly
plugged into the server computer and the telephone line. If
necessary, use the VoiceCallTester tool provided by Thermo
Scientific.
For a dry contact device: make sure the device is properly connected
to its power supply and the network.
No communication with Smart Service [-995]
Check the application logs to see why the service stopped.
After correcting the problem, restart the Smart-Svc for the receiver
in question using the Windows Services manager.
If the problem continues, restart the computer (or at least
Smart-Vue Server).
No communication with receiver [-996]
USB receiver plugged in locally:
Restart the Smart-Service, if failed; reboot computer.