TCP/IP dry contact device
Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution Administrator Guide v 2.2 89
5. Click on OK to save your network settings.
6. Click on Apply Settings in the left-hand panel to save your
configuration changes to the network receiver. You may close your
browser after the configuration is saved.
7. Re-start the receiver by unplugging the AC adapter and plugging it back
in.
17.7 TCP/IP Dry
Contact Software
configuration
For details on using Smart-Vue Client to add a dry contact to your
monitoring system, please see the Smart-Vue Software User Manual.
17.7.1 Testing the TCP/IP Dry
Contact device
This section describes how to run a sample test after adding the dry contact
to your system as described in Software User Manual. The test procedure
described here generates a false alert with a time/date stamp of 12/30/1899
at 00:00 so you can identify these events as tests easily. Also, they are not
recorded in the alarm log.
To test your dry contact, along with all other alerts you may have configured
(fax, SMS, e-mail, etc.), do the following. Enter your login and password
when prompted.
1. In Smart-Vue Client, select the Tools tab and then click on Alert test
().
2. Enter your login name and password, if necessary, to continue.