Wireless Siren
80 Administrator Guide v 2.2 Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution
16 Wireless Siren
16.1 Package Contents Smart-Vue wireless siren
AC power adapter with 1 meter (3.3 ft.) cable and 4 international plug
sizes; 100/240 V (50/60 Hz) AC 12 V DC – 1 A
Mounting kit with 2-sided Velcro® tape
Documentation CD-ROM
16.2 Siren Module
Wireless Specifications
ISM (Industrial Scientific Medical) band with four frequencies:
US/CAN 915 MHz; Europe 868 MHz; APAC 434 MHz; India and
other countries 867 MHz
FCC 15 compliant; CE EN-300-220
Channel width: 50 kHz
Frequency deflection: 16 KHz
Transmission speed: 9600 Baud in NRZ mode
Modulation type: GFSK
Driven receiver sensitivity for BER = 1%: -107dbm to -110 dBm
Driven transmission: 8 dBm to 10 dBm
Power output: 25 mW (867, 868 and 915 MHz), 10 mW for 434
MHz.
Plastic enclosure: ABS and polycarbonate with IP54 protection
Wireless module operating conditions: 0°C to +50°C (+32° to +122°F);
0 to 90% RH, non-condensing
Size: 160 x 80 x 83 mm (1.6 x 3.1 x 3.3 in.)
Weight: 400 g (14.1 oz.)
16.3 Product Overview The Smart-Vue wireless siren is an AC-powered wireless module with siren
and flashing light. It is easy to set up in Smart-Vue Client software.
16.3.1 Main Features Siren volume output from 60 to 100 dBA (+/- 10%)
Wireless range up to 200 meters (650 ft) in open space, from 8 meters
(26 ft) to 30 meters (98 ft) indoors, and up to 120 meters (390 ft)
indoors with 3 repeaters