Preventive Maintenance Guidelines
136 Administrator Guide v 2.2 Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution
• Alarm caused by an absence of communication with the
Smart-Service for the receiver in question: see instructions for error
code -995 (later in this document).
• Make sure that no alerts are related to technical alarms on alert
devices (or labeled "Media error").
• Absence of communication with the telephone modem: see
instructions for error code -994 (later in this document).
• Network not available: see instructions for error code -994.
• Absence of communication with the wireless siren: see instructions
for error code -994 (later in this document).
• Check the alarm list to see which alarms must be acknowledged, then
acknowledge the alarms for which the problem has been resolved.
• We strongly recommend that you acknowledge alarms on a regular basis
and not allow them to accumulate for too long. Too many pending
alarms can slow down the system and interfere with client application
connections.
• Back-up your data every day
• Contact your database administrator to setup a daily backup
strategy.
• By default, the database is installed in:
C:\SmartVue > Mysql > mysql server 5.0 > data > thermo
22.2.3 Weekly • Perform an “Alert Test” before the weekend (Smart-Vue Client Tools
Tests Alert test)
• This enables you to test your configuration and related alert
equipment.