TCP/IP Receiver
48 Administrator Guide v 2.2 Thermo Scientific Smart-Vue
®
Wireless Monitoring Solution
12.3.1 General Requirements
and Recommendations
The machine running Smart-Vue Server software runs continuously,
24/7/365, and should be connected to an Uninterruptible Power Supply
(UPS) that protects against power surges and provides power to back up
the machine and components.
The AC adapter used for your network receiver should also be plugged
into the UPS.
The UPS is also recommended as a power backup for your
communication/network systems
Weekly manual system testing (e.g., pull a sensor, verify that the
communication system is working) should be performed as defined in
the Standard Operating Procedure (SOP).
Recommended maintenance and calibration procedures should be
followed.
If you are storing cold products, Thermo Scientific recommends use of a
back-up cooling system (e.g., CO
2
or LN
2
) to maintain freezer chamber
temperature below the critical level should a power failure occur.
Contact your local sales representative for more information.
Thermo Scientific recommends Installation and Operational
Qualifications (IQ/OQ) be performed before initial use.
For timely notification of an emergency/alarm, it is critical to establish
primary, secondary and tertiary call-out procedures with escalation so
contact is not dependent on reaching just one individual. Ideally,
ultimate contact is with a security department or company with 24/7
monitoring.
Please see the Smart-Vue Client 2.2 Software User Manual for an in-depth
description of Smart-Vue end-point module configuration, alerts,
notification, data management and general system features.
12.3.2 System Requirements
CAUTION: Do not install Smart-Vue Server on a laptop computer.
For Windows system compatibility, refer to Section 21.2 “Software
compatibility”.
Machine running the following Microsoft Windows 32- or 64-bit
systems:
Windows 7, Windows 8, Windows 10, Windows Server 2008,
Windows Server 2012, and Windows Server 2016
10/100/1000 Mbit Ethernet network (TCP/IP)

Table of Contents

Other manuals for Thermo Scientific Smart-Vue

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Smart-Vue and is the answer not in the manual?

Thermo Scientific Smart-Vue Specifications

General IconGeneral
BrandThermo Scientific
ModelSmart-Vue
CategoryControl Unit
LanguageEnglish